Michigan Instruments, Inc. :: Calibration Info
   home  •  contact us  

CALIBRATION INFO
Customer Support


Factory Calibration

Factory calibration encompasses all aspects of sensor, electronic, mechanical and software tuning to ensure maximum accuracy of all measurements and simulated parameters provided by TTL/PneuView systems.

Calibration procedures include setting of the compliance and resistance characteristics of the lungs and setting offset and gain characteristics for each of the pressure transducer channels.

Multiple data points are taken at different compliance settings, which are mathematically analyzed and then used to produce a Compliance Characteristic Calibration Table individually for each lung. It is these calibration tables that the volume calculation algorithms described in the previous section use to determine the coefficients to use for the real time solution of the dynamic equations.
 
Calibration Verification by the User
Pressure and Volume calculation accuracy may be easily verified with the aid of a calibrated syringe and independent pressure measurement device. All versions of the PneuView system ship with pre-formatted templates to aid the user with the verification of calibration accuracy. These templates include step-by-step instructions, which should be followed periodically to ensure maintenance of calibration accuracy or anytime accuracy comes  into question, such as if the unit suffers some type of physical damage.
Below is an example of a Certification of Calibration


Click to enlarge

 

Recalibration Service

To reduce the potential for an inaccurate reading due to the passage of time or extended use, a recalibration is recommended for your instrument every 2 years for Models 3600i, 5600i & 5601i
and every 3 years for models 4600, 1600 & 1601.

To schedule your return please complete the following:

Company Name:
Contact Name:
Address:
City:
State:
Postal Code:
Phone Number:
E-Mail Address:
Fax Number:
 
TTL Model Number:
Serial Number:
Calibration:

If you would like "As Received" data taken, please check this box (additional charge)

 
Any Additional Service Requirements, Questions or Comments:

 

Ship to:

Reference RA
Michigan Instruments
4717 Talon Ct SE
Grand Rapids, MI 49512-5409

   

Clean and decontaminate your equipment so it is safe to handle.

Package the item well to avoid damage during shipping.

NOTICE: Please DO NOT ship any accessories or miscellaneous items back
with the unit. We will not be held responsible for any items other than the primary unit.

 

If you have any questions or concerns, please contact us at (616) 554-9696 x309 or Click Here to contact us via e-mail. Our service department will be contacting you within 24 business hours with a Return Authorization #.


 
 